Human Renin Quantikine ELISA Kit Summary

貨號:DREN00 

Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (25 uL), EDTA Plasma (25 uL), Heparin Plasma (25 uL), Urine (50 uL)
Sensitivity
14.8 pg/mL
Assay Range
31.3 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine)
Specificity
Natural and recombinant human Renin. This assay recognizes both the Pro- and mature versions of recombinant human Renin.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.

Product Summary

The Quantikine Human Renin Immunoassay is a 4.5 hour solid-phase ELISA designed to measure human mature Renin and Prorenin in cell culture supernates, serum, plasma, and urine. It contains NS0-expressed recombinant human Renin and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human Renin showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for naturally occurring human Renin.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision

Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean 201 599 1193 191 566 1153
Standard Deviation 10.6 14.2 19.9 10.5 27.1 46.4
CV% 5.3 2.4 1.7 5.5 4.8 4

Recovery

The recovery of Renin spiked to levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=8) 100 93-106
EDTA Plasma (n=4) 101 95-107
Heparin Plasma (n=4) 105 99-114
Serum (n=4) 101 90-109
Urine (n=4) 93 80-106

Linearity

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Renin were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Human Renin ELISA Linearity